V118 Single-use vials punctured only once
Written V0118 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
CDC RR-05 as Adopted by Reference 42 CFR 494.30(a)(1)(i)
Intravenous medication vials labeled for single use, including erythropoietin, should not be punctured more than once.
CMS's interpretive guidance
Interpretive Guidance § 494.30(a)(1)(i)
According to the CDC, once a needle has entered a vial labeled for single use, the sterility of the product can no longer be guaranteed. Residual medication from two or more vials should not be pooled into a single vial.
Single-use vials/ampules must be used for only one patient, should not be entered more than once, and if entered, may not be stored for future use.
Staff should only enter vials with a new sterile syringe and needle. If both vials are single-use and are discarded after the single entry into each, the same syringe may be used. If either vial is multi-use, a different syringe must be used for entry into each vial.