V120 External transducer protectors
Written V0120 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
CDC RR-05 as Adopted by Reference 42 CFR 494.30(a)(1)(i)
Use external venous and arterial pressure transducer filters/protectors for each patient treatment to prevent blood contamination of the dialysis machines’ pressure monitors.
If the external transducer protector becomes wet, replace immediately and inspect the protector. If fluid is visible on the side of the transducer protector that faces the machine, have qualified personnel open the machine after the treatment is completed and check for contamination. This includes inspection for possible blood contamination of the internal pressure tubing set and pressure sensing port. If contamination has occurred, the machine must be taken out of service and disinfected using either 1:100 dilution of bleach (300–600 mg/L free chlorine) or a commercially available, EPA-registered tuberculocidal germicide before reuse.
Change filters/protectors between each patient treatment, and do not reuse them. Internal transducer filters do not need to be changed routinely between patients.
CMS's interpretive guidance
Interpretive Guidance § 494.30(a)(1)(i)
According to the CDC, external transducer protectors, which provide a protective barrier between dialysis bloodlines and the dialysis machine, should not be reused. “Wet” (“wet with blood or other fluid”) external transducer protectors must be changed immediately, and the side of the external transducer protector that faces the machine should be inspected for visible fluid. If the external transducers are wetted with blood, the staff should inspect the wetted transducer to see if fluid has passed through. If fluid or blood is visible on the side of the transducer protector that faces the machine, the machine must be opened by qualified personnel after the dialysis treatment to allow the internal transducer to be inspected for contamination. This includes inspecting the internal pressure tubing set and pressure sensing port for possible blood contamination. Frequent blood pressure alarms or frequent requirements for adjusting the blood level in the drip chamber can be indicators of contamination of the internal transducer filter.
Recognize that some bloodlines do not have external transducer protectors; this requirement would not apply in those cases, except for changing the bloodlines between patients.