V122 Cleaning and disinfecting surfaces and equipment

Written V0122 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.30(a)(4)(ii) · Appendix H Rev. 246, page 16 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§494.30(a)(4) - And maintaining procedures, in accordance with applicable State and local laws and accepted public health procedures, for the—
(ii) - Cleaning and disinfection of contaminated surfaces, medical devices, and equipment.

Appendix H quotes this text a little differently from the regulation as published today. The eCFR has the current wording.

Read 42 CFR 494.30 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.30(a)(4)(ii)

Failures in environmental cleaning and disinfection have led to transmission of blood-borne pathogens (e.g., Hepatitis B virus) and other infections from one patient to another in hemodialysis units. Correct cleaning and disinfection of environmental surfaces (including patient chair or bed surfaces, dialysis equipment surfaces, adjacent tables, and work surfaces) must be performed between patient uses to prevent transmission of dangerous pathogens.

A facility should establish written protocols for cleaning and disinfecting surfaces and equipment, including careful mechanical cleaning before any disinfection process. For guidance, refer to CDC RR-5 Table 2 below.

CDC Table 2. Disinfection procedures recommended for commonly used items or surfaces in hemodialysis units

Any manufacturer’s guidance for sterilizing or disinfecting an item should be followed, as well as guidance from the chemical sterilant or disinfectant manufacturer, including appropriate dilution and contact time.

At the end of each dialysis treatment, all surfaces without visible blood should be cleaned following the low-level disinfection protocol using soap, detergent or detergent germicide. Per CDC guidance, for visible blood, the intermediate-level disinfection protocol must be followed, which requires the area to be immediately cleaned with a cloth soaked in tuberculocidal disinfectant or a 1:100 dilution of bleach (300-600 mg/L free chlorine), following the manufacturer’s directions for contact time. Gloves must be worn, and the used cloth placed into a leak-proof container. After cleaning up all visible blood, a disinfectant must be applied a second time using a new cloth or towel. No patient should be at the station during this time.

Blood spills in the treatment area and other areas, such as the waiting room and patient bathroom, need to be cleaned effectively and immediately, or as soon as possible, given the patient care situation. Cleaning and disinfection of a blood spill must be completed with a cloth soaked with a tuberculocidal disinfectant or a 1:100 or stronger dilution of bleach (300-600 mg/L free chlorine) (i.e., intermediate-level disinfection). After all visible blood is cleaned, staff should use a new cloth or towel to apply disinfectant a second time.

In hemodialysis units, cleaning and disinfection procedures during patient changeover are particularly prone to error and contribute to the risk of cross-contamination if correct procedures are not observed. A sufficient patient-free interval is necessary at each station to facilitate adequate cleaning and disinfection. Routine surface disinfection should not begin until the patient has left the station. Station disinfection includes the following equipment/surfaces: the dialysis machine, the purified water connection, dialysate concentrate container(s) or connection(s), the wall-box, and the treatment chair.

According to current recommendations from the CDC, to prevent cross-contamination between patients, it is essential that the previous patient completely vacate the station before staff begin cleaning and disinfecting the station and setting up for the next patient. It is essential to note that all dialysis patients must be clinically stabilized (i.e., have stable blood pressure and vascular access hemostasis) following their dialysis treatments before being transferred from the dialysis station. If a patient is not sufficiently stable to be moved from the dialysis station, cleaning and disinfection of the equipment at the station, as well as preparations for the next patient, must be delayed until the patient can be safely moved outside the station. CDC has developed an audit tool and checklist to promote recommended practices for infection prevention in dialysis facilities. They are available for use by facility staff to help guide and assess their practice. These resources can be found here:

“Intermediate-level disinfection” kills bacteria and most viruses and is accomplished by using a tuberculocidal “hospital disinfectant” or a 1:100 dilution of bleach (300-600 mg/L free chlorine). “Low-level disinfection” kills most bacteria and is accomplished by using general-purpose disinfectants. Intermediate and low-level disinfectants are designed for use on environmental surfaces; they can also be used on non-critical medical devices, depending on the design and labeling claims. Reference https://www.cdc.gov/mmwr/pdf/rr/rr5005.pdf, page 15/46.

At the end of each patient treatment, the staff should clean and disinfect the dialysis station. Special attention should be given to cleaning control panels on dialysis machines, treatment chairs, and other frequently touched surfaces that may be contaminated with patients’ blood. The staff should discard all fluids and clean and disinfect all surfaces of the containers associated with the prime waste (including containers attached to the machines) after each treatment.

After each treatment, the staff needs to clean and disinfect medical devices and equipment. Items such as scissors, hemostats, clamps, stethoscopes, and blood pressure cuffs need to be cleaned and disinfected between patient uses. If the item is visibly contaminated with blood, an intermediate-level disinfectant must be used.

Staff must clean and disinfect the internal circuits of the dialysis machines properly. Single-pass machines may be rinsed and disinfected at the beginning or end of each day, while batch-recirculating machines must be drained, rinsed, and disinfected after each use. If a blood leak occurs, follow the manufacturer’s recommendations for additional disinfection.

A facility should document procedures for the dialysis machine disinfection, including testing for residual disinfectant.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 16 (the full 469-page PDF) before quoting it.