V175 Water and dialysate quality (the Condition)
Written V0175 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
§494.40 - Condition: Water and dialysate quality.
CMS's interpretive guidance
Interpretive Guidance § 494.40
This Condition incorporates the Association for the Advancement of Medical Instrumentation’s (AAMI’s) “American National Standard for Dialysate for Hemodialysis,” 2004 (RD52:2004) into the regulation. This AAMI document references portions of their “American National Standard for Water Treatment Equipment for Hemodialysis Applications” (RD62:2001) as the specifications for various water treatment components. The referenced portions of RD62:2001 are also incorporated into the regulation by reference.
The V-tags associated with the water and dialysate quality requirements are divided into 3 parts: (1) the 2004 AAMI standard, (2) any related 2004 AAMI guidance that CMS identifies as relevant, and (3) “Additional Guidance” where CMS provides its additional suggestions on how to best comply with the AAMI guidance.
A survey of this Condition requires inspection of the water treatment and dialysate preparation equipment, as well as their distribution systems; an interview with personnel responsible for the day-to-day operation of those systems; and a review of the records of operation and testing for those systems. Supervisory personnel may be interviewed to clarify issues or questions. Critical to ensuring patient safety is the expectation that every survey visit include direct observation of water testing for chlorine/chloramine.
Noncompliance at the Condition level should be considered if identified deficient practices are pervasive throughout the Standards included in this Condition, serious in nature, and/or a potential risk to patient health and safety.
Examples of potential Condition level noncompliance may include, but are not limited to:
- Demonstrated lack of knowledge or training of staff assigned responsibility for the operation or monitoring of the water treatment or dialysate preparation systems;
- Failure to perform and document the test(s) for chlorine and chloramine accurately, including the use of testing strips or reagents that are expired or not sensitive to the required levels;
- Unsafe practices in the preparation, labeling, or delivery of dialysate; and,
- Failure to address out-of-range results for tests of water or dialysate (bacteria, endotoxin, or chemical analysis).