V234 Bicarbonate concentrate: avoiding over-mixing

Written V0234 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 101 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.4.4.3 Bicarbonate concentrate mixing systems

Over-agitating or over-mixing of bicarbonate concentrate should be avoided, as this can cause CO2 loss and can increase pH.

ANSI/AAMI RD52:2004

5.4.4.3 Bicarbonate concentrate mixing systems

Systems designed for mixing dry acid concentrates may use methods that are too vigorous for dissolving dry bicarbonate.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

AAMI Rationale for the Development and Provision of This Recommended Practice

5.4.4.3 Bicarbonate concentrate mixing systems

Over-agitation or mixing of bicarbonate concentrate may result in loss of CO2 from the solution. Loss of CO2 results in an increase in pH and favors the formation of carbonate that can lead to precipitation of calcium and magnesium carbonate in the fluid pathways of the dialysis machine following dialysate proportioning.

Additional Guidance:

There should be a system to prevent over-mixing of bicarbonate. This could include a timer integrated into the mixing system for automatic cut-off, or a policy to require staff to monitor the mixer and cut it off immediately when the time period for mixing is completed. Use of over-mixed bicarbonate concentrate can result in a low calcium level in the dialysate and a concomitant drop in patients’ serum calcium levels.