V244 Bicarbonate jugs: disinfection at least weekly

Written V0244 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 108 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.5.4 Bicarbonate concentrate distribution systems (disinfection of bicarbonate jugs)

6.5 Concentrate distribution 7 Strategies for bacterial control

7.1 General

5.5.4 Bicarbonate concentrate distribution systems (disinfection of bicarbonate jugs)

When reusable concentrate jugs are used to distribute bicarbonate concentrate, they should be rinsed free of residual concentrate before disinfection.

6.5 Concentrate distribution

When reusable concentrate jugs are used to distribute bicarbonate concentrate, they should be disinfected at least weekly.

7 Strategies for bacterial control

7.1 General

Following disinfection, jugs should be drained, rinsed, and inverted to dry.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004 7 Strategies for bacterial control

7.1 General

Facilities that reuse concentrate jugs for bicarbonate concentrate should disinfect the jugs at least weekly. Bicarbonate concentrate can support prolific growth of microorganisms. Jugs can be disinfected with household bleach solutions (300 mg to 600 mg free chlorine, or 30 mL to 60 mL of 6.15% household bleach per gallon of water) with a contact time of about 30 minutes or another EPA-registered disinfectant according to the manufacturer’s instructions.

Additional Guidance:

A 1:100 solution of household bleach: treated water may be used, which yields about 500-615 ppm available chloride. If the facility uses less contact time than 30 minutes, there must be evidence of the use of dialysate culture results to determine the needed disinfectant contact time. In all cases, the bicarbonate container disinfection process must result in compliance with the allowable dialysate microbiological levels of <200 cfu and <2 EU.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 108 (the full 469-page PDF) before quoting it.