Middle Molecule Find

Rulebook › V-tags › V278

V278 Preconfigured HD systems: quarterly bacteria and endotoxin

Written V0278 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.40(e) · Appendix H Rev. 246, page 129 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.40(e) - …Moreover, the facility must perform bacteriological and endotoxin testing on a quarterly, or more frequent basis, as needed, to ensure that the water and dialysate are within AAMI limits.

Read 42 CFR 494.40 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.40(e)

Preconfigured systems used for in-center treatments must be tested at least quarterly for bacteria and endotoxins. Testing should occur more frequently when needed.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 129 (the full 469-page PDF) before quoting it.

Connected tags

Tags whose guidance points here