Written V0304 on the CMS-2567 and in QCOR.

Enforces 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 131 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

§ 494.50(b) - Reprocessing requirements for the reuse of hemodialyzers and bloodlines: A dialysis facility that reuses hemodialyzers and bloodlines must adhere to the following reprocessing guidelines:

(1) Meet the requirements of AAMI published in “Reuse of Hemodialyzers,” third edition, ANSI/AAMI RD47:2002 and RD47:2002/A1:2003. The Director of the Federal Register approves this incorporation by reference in accordance with 5 U.S.C. 552(a) and 1 CFR Part 51. This publication is available for inspection at the CMS Information Resource Center, 7500 Security Boulevard, Central Building, Baltimore, MD or at the National Archives and Records Administration (NARA). For information on the availability of this material at NARA, call 202-741-6030, or go to: http://code_of_regulations/ibr_locations.html. Copies may be purchased from the Association for the Advancement of Medical Instrumentation, 3300 Washington Boulevard, Suite 400, Arlington, VA 22201- 4598.

Appendix H quotes this text a little differently from the regulation as published today. The eCFR has the current wording.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)

The AAMI “Reuse of Hemodialyzers,” third edition, ANSI/AAMI RD47:2002/A1:2003 is incorporated by reference. The recommendations are provided in the “Regulation” column and carry the full weight of regulation. Explanatory language from the AAMI recommendations and portions of the AAMI rationale for the development and provisions of the recommended practice are provided in the “Interpretive Guidance" column as an aid to understanding the regulation. In some cases, the AAMI identification numbers may appear out of order, as they have been rearranged for better flow with the survey process. While exact language from the AAMI document has been incorporated as regulation, the language in the “Interpretive Guidance” column has been edited for clarity, brevity, and to minimize redundancy.

Note: When words are inserted or altered in a direct quotation, square brackets—[ ]— are placed around the change. The brackets enclose words intended to explain the quote or to help integrate the quote into the guidance.

AAMI RD47:2002/A1:2003 11 Reprocessing

The multiple use of a dialyzer begins with the labeling of the new dialyzer (see AAMI section 10) and then continues with the reprocessing procedures described in this section. Preparation of the reprocessed dialyzer for the next dialysis is described in AAMI section 12. The cycle is repeated after the next use of the dialyzer until the dialyzer does not meet the criteria for continued use. The results of the tests and the signature or other unique means of identifying the person performing each step should be maintained in a permanent record (see AAMI section 4.2). Completion of all reprocessing steps, tests, and inspections should be documented in the reprocessing record, accompanied by the signature or other unique means of identification of the person completing them. When appropriate for the reprocessing procedure in use, all dialyzer manufacturer’s instructions regarding reuse should be carefully followed.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 131 (the full 469-page PDF) before quoting it.