V306 Dialyzer reprocessing manual

Written V0306 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 132 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

4.1 Dialyzer reprocessing manual

The dialyzer reprocessing manual should be a compilation of all specifications, policies, training materials, manuals, methodologies, and procedures that may be integrated into the dialysis facility’s policy and procedures manual. The dialyzer reprocessing manual should also contain samples of forms and labels, if appropriate. The operational logs, manuals, and files may be kept separate from the dialyzer reprocessing manual. The dialyzer manufacturer’s labeling should be consulted to determine if a specific dialyzer requires special considerations.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.4 Records

Documentation is essential to a safe, effective hemodialyzer reprocessing program. The overall dialyzer reuse procedure documentation includes reference materials, procedures, and policies, some of which may be distributed in the facility for operating purposes. The other records serve to document aspects of the reuse procedure for each dialyzer, along with QC and QA measures, so that a complete history of the reprocessing of each dialyzer and QC/QA procedures exists.

Allowance is made for keeping the reprocessing record data in the reprocessing log, the patient’s chart, or a combination of the two, because both of them are traceable, permanent records, and it may be inconvenient to record all of the information in one location.

Additional Guidance:

The reprocessing manual must be complete for the reprocessing method, germicide, and system in use. The manual must address test procedures, maintenance and calibration of the reprocessing equipment and training and competency testing of personnel. The manual may be separate or combined with the general policy and procedure manual. The manufacturer’s “labeling” of a dialyzer includes the package insert, usually found in the shipping container for multiple dialyzers of the same type, and contains the dialyzer manufacturer’s directions for use. If a specific dialyzer in use requires special consideration, the reprocessing manual should reflect the manufacturer’s guidance.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 132 (the full 469-page PDF) before quoting it.