V325 Germicide concentration test and heat disinfection proof

Written V0325 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 147 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

7.2.2 Process control testing: concentration of germicide

7.2.2.2

The test for the concentration of germicide or chemical shall be established before clinical use of the reprocessed dialyzers ([AAMI] 11.4.1.6 and 12.3.2). For systems using heat disinfection, verifiable evidence shall be available before the next use that dialyzers have been exposed to the appropriate temperature for the time required. If chemicals are used to enhance heat disinfection, both a presence test and a verification of time and temperature shall be performed.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

The reuse manual should document how the concentration of germicide will be tested.

If heat disinfection is the reprocessing method, records of each batch of dialyzers processed must include an indicator, such as an automated time/temperature recording log, that the dialyzers were exposed to the appropriate temperature for the time required. If a chemical, such as citric acid, is used to enhance heat disinfection, a presence test is also required before clinical use of the dialyzers.

If an incubator or oven is used to raise the dialyzer storage temperatures, a recording thermometer should be in use to ensure sufficient temperature is consistently maintained. Records should document that these devices functioned as expected.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 147 (the full 469-page PDF) before quoting it.