Middle Molecule Find

Rulebook › V-tags › V327

V327 One reprocessed dialyzer, one patient

Written V0327 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 148 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

10 Hemodialyzer labeling

Each reprocessed hemodialyzer shall be used for only one patient. The labeling shall uniquely identify the patient who is using the dialyzer. The dialyzer should also be labeled with other information essential to proper reuse procedure.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

Precautions must be in place to ensure that each dialyzer is used only for one patient. Each reprocessed dialyzer must have a permanently affixed label uniquely identifying the patient using that dialyzer. Each patient must always receive treatment on their own dialyzer, and dialyzers must not be mismatched to patients.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 148 (the full 469-page PDF) before quoting it.