V378 Reprocessing method: manufacturer's or validated alternate
Written V0378 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
§ 494.50(b)(2) - Reprocess hemodialyzers and bloodlines – (i) By following manufacturer’s recommendations; or (ii) Using an alternate method and maintaining documented evidence that the method is safe and effective.
Appendix H quotes this text a little differently from the regulation as published today. The eCFR has the current wording.
CMS's interpretive guidance
Interpretive Guidance § 494.50(b)(2)
Each manufacturer of dialyzers for multiple use is required by FDA to provide at least one acceptable reprocessing method with at least one germicide. The facility may use that method/germicide or choose an alternate method. If the facility has chosen an alternate method, there must be documentation that the chosen method has been validated as safe and effective.
https://middlemolecule.com/rulebook/tags/v378/