V635 QAPI area: dialyzer reuse, if the facility reuses
Written V0635 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
§ 494.110 (a)(2) – The program must include, but not be limited to, the following: (vii) Hemodialyzer reuse program, if the facility reuses hemodialyzers.
CMS's interpretive guidance
Interpretive Guidance § 494.110 (a)(2)(vii):
If a facility has a dialyzer reuse program, it must comply with the quality assurance requirements specific to reuse, located at V360-368. These requirements outline periodic reuse processes and practice audits, which must be conducted and documented to ensure the reuse program remains safe and effective. Refer to the Measures Assessment Tool (MAT) for quality indicators related to the reuse of hemodialyzers.
The QAPI meeting minutes should demonstrate oversight of the reprocessing/reuse program and include at least summaries of the required reuse audits.
This tag is used to cite major problems related to quality assessment and performance improvement for the Condition for reuse. Use individual tags in the Condition for reuse of hemodialyzers for citations of specific or isolated issues in the required audits for quality assurance (V360-V368).
Connected tags
V360 to V368, the range named above
- V360 Reuse quality assurance
- V361 Tracking problems until the fix works
- V362 Annual audit: informing patients about reuse
- V363 Annual audit: equipment manuals and procedures
- V364 Audits: reprocessing area, storage, protection
- V365 Semiannual audit: reprocessing supplies
- V366 Quarterly audit: dialyzer labeling
- V367 Audit: reprocessing procedures
- V368 Quarterly audit: preparation for dialysis
Tags whose guidance points here