V640 Immediate correction of threats to patients
Written V0640 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
§ 494.110(c) – Standard: Prioritizing improvement activities.…The facility must immediately correct any identified problems that threaten the health and safety of patients.
CMS's interpretive guidance
Interpretive Guidance § 494.110(c):
Examples of conditions which could pose a threat to the health and safety of dialysis patients and require immediate correction include, but are not limited to:
- Dangerous levels of contaminants in product water;
- Unsafe levels of electrolytes in dialysate;
- Failure to conduct an accurate pre-assessment;
- Setting an inaccurate fluid removal rate;
- Failure to provide adequate observation of patient, patient vascular access, patient equipment;
- Defective clinical equipment;
- Failure to adequately disinfect reprocessed dialyzers; Failure to reduce residual germicides in reprocessed dialyzers to safe levels;
- Lack of qualified staff to perform crucial tests or to meet critical patient needs;
- Evidence that staff assigned to perform crucial tests or to meet critical patient needs are not competent;
- Potential for cross-contamination between infected and non-infected patients; and,
- Failure to use machine-provided safety devices (muting machine alarms, bypassing the air detector, etc.).
The facility must take immediate, appropriate actions to address any serious threats and ensure patient health and safety.