V182 Contingency plans for water system failure
Written V0182 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5 Equipment
5.1 General:
A dialysis facility should develop contingency plans to cover the failure of its water purification and distribution system or a critical component of that system. Such contingency plans should describe how to deal with events that completely prevent dialysis from being performed, such as failure of the facility’s municipal water supply or electrical service following a natural disaster or water main break. Other plans should address how to deal with sudden changes in municipal water quality, as well as with the failure of a critical component of the water purification and distribution system.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5 Equipment
5.1 General
This section on equipment provides a brief description of the different components that may be included in a water purification and distribution system used for hemodialysis applications. Since feed water quality and product water requirements may vary from facility to facility, not all of the components described will be necessary in every purification and distribution system.
Routine dialysis requires a well-functioning water purification and distribution system, since dialysis cannot be performed without an adequate supply of water. In addition, specific components of the water purification and distribution system are critical to its operation. An example of such a critical component is the circulating pump in an indirect feed system.
Additional Guidance:
An emergency or contingency “plan” that specifies the facility will use tap water or de-chlorinated tap water may include evidence that the water intended for use has been found safe for such use (i.e., has levels below AAMI accepted limits of aluminum, copper, chloramines, fluoride, nitrate, sulfate, zinc and other contaminates known to be toxic to dialysis patients). Refer also to the requirements for emergency policies and procedures found under the Condition for Physical Environment at V408.