V177 Water: maximum chemical contaminants (Table 1)

Written V0177 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 37 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

4 Fluid quality

4.1 Water

4.1.1 Maximum level of chemical contaminants in water

Product water used to prepare dialysate or concentrates from powder at a dialysis facility, or to process dialyzers for reuse, shall not contain chemical contaminants at concentrations in excess of those listed in ANSI/AAMI RD62 … which is reproduced in Table 1 below.

The manufacturer or supplier of a complete water treatment system should recommend a system that is capable of meeting the requirements of this clause at the time of installation given the analysis of the feed water. The system design should reflect possible seasonal variations in feed water quality.

Following installation of a water treatment, storage, and distribution system, the user is responsible for continued monitoring of the levels of chemical contaminants in the water and for complying with the requirements of this standard.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52: 2004 4 Fluid quality

4.1 Water

The fluid quality requirements apply to the purified water as it enters the equipment used to prepare dialysate or concentrates from powder at a dialysis facility and apply to the water treatment system as a whole and collectively, and not to each of the devices that make up the system.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.4 Fluid quality

A.4.1.1 Maximum level of chemical contaminants in water

ANSI/AAMI RD62:2001 establishes maximum levels of chemical contaminants in three categories: chemicals known to have particular toxicity for hemodialysis patients, chemicals listed in the U.S. Environmental Protection Agency’s Safe Drinking Water Act, and physiological substances that can adversely affect patients if present in the dialysate in excessive amounts.

Several chemicals have been clearly shown to be toxic to dialysis patients at concentrations that are not necessarily toxic to the general population. Those chemicals include aluminum, copper, chloramines, fluoride, nitrate, sulfate, and zinc.

Uptake of aluminum from the dialysate is associated with bone disease, anemia, and dialysis encephalopathy syndrome, which is usually fatal. The suggested maximum aluminum level has been specified to prevent the accumulation of this toxic metal in the patient. Aluminum is particularly likely to increase suddenly to high levels as a result of changing the method of water treatment to include aluminum-containing compounds.

Chloramines damage red blood cells by oxidizing hemoglobin to methemoglobin and by inhibiting antioxidant pathways. Their toxicity in hemodialysis patients is undisputed. Although the role of free chlorine in oxidative blood damage is unclear, its high oxidation potential and ability to form chloramines suggest that the use of highly chlorinated water in dialysate preparation should be avoided.

High levels (>20 ppm) of fluoride in the water used to prepare dialysate are clearly toxic to hemodialysis patients and have resulted in patient deaths. Such high levels of fluoride have resulted from accidental over fluoridation of a municipal water supply, as well as from deionizer exhaustion. The toxicity of fluoride in dialysis patients is questionable at the levels usually associated with fluoridated water (1 ppm). However, in the absence of a consensus on its role in uremic bone disease, the AAMI Renal Disease and Detoxification (RDD) Committee thought it prudent to restrict the fluoride level of dialysate.

Nitrates are a marker for bacterial contamination and fertilizer runoff and have been linked to cases of methemoglobinemia. Nitrates should be permitted only at very low levels. Sulfate at levels above 200 mg/L has been related to nausea, vomiting, and metabolic acidosis. The symptoms disappear when the level remains below 100 mg/L. Both copper and zinc toxicity have been demonstrated when those substances are present in dialysate at levels below those permitted by the EPA standard. Hence, a lower level has been chosen.

The second group of chemical contaminants included in ANSI/AAMI RD62:2001 is based on the U.S. Environmental Protection Agency’s Safe Drinking Water Act (see 2.6). The standard specifies maximum allowable limits for most contaminants in this group at 1/10 of the EPA maximum allowable limit. The lower levels were chosen because the volume of water used for dialysis far exceeds that used for drinking water, because protein binding of these solutes may occur in the blood, and because there is reduced renal excretion of these substances. Selenium and chromium levels were set at the “no-transfer” level. The no-transfer level was chosen, even though it exceeds the EPA limit for selenium and 28% of the EPA limit for chromium, because there is no need for a restriction below the level at which there is no passage from the dialysate to the blood.

The third group of substances included in ANSI/AAMI RD62:2001 consists of physiological substances that can adversely affect the patient if they are present in the dialysate in excessive amounts. Calcium, potassium, and sodium are examples of those substances.

The chemical contaminants regulated by ANSI/AAMI RD62:2001, as listed in Table 1, should not be considered a definitive list of harmful substances; they are only a partial listing of contaminants that might reasonably be expected to be present and have clinical implications. Iron is not included because it does not enter the patient’s blood in sufficient quantities to cause toxicity. However, iron may cause fouling of water purification devices or dialysate proportioning systems. Furthermore, municipal water supplies are dynamic systems, which may change with the seasons or in response to new regulations from the EPA.

Additional Guidance:

Table 1 reflects the Standard adopted by CMS as regulation on April 15, 2008, 73FR20369. Individuals or groups of facilities may adopt newer, more stringent requirements by policy.

The medical director is ultimately responsible for the safety and quality of the water used for patient treatments. Each product water chemical analysis must be within the parameters listed in Table 1 at V177. If any values exceed those listed, facility staff must notify the medical director of the results, repeat testing, and take action to address any repeated high levels.

The medical director must be knowledgeable about the water treatment system installed and ensure that the system, as installed, will produce AAMI quality water. Ways to ensure this result would include analysis of the source water and consultation with experts in water treatment, as well as confirmation that the planned installation would be sufficient to produce AAMI quality water by the manufacturer or vendor of the water treatment equipment. For initial surveys, the facility should provide a copy of a chemical analysis with results within AAMI standards accomplished before starting any patient treatment in the new facility. For resurveys, there must be evidence of ongoing monitoring of the chemical quality of the water, along with actions taken when levels exceed the AAMI standards.

The use of water outside of AAMI standards should be infrequent, considered only when no other option is available to provide desperately needed dialysis, and limited to one treatment per patient. An emergency “plan” that specifies the facility will use tap water or dechlorinated tap water is not acceptable without evidence that the source water has been found safe for such use (i.e., has levels below AAMI accepted limits of aluminum, copper, chloramines, fluoride, nitrate, sulfate, zinc, and other contaminants known to be toxic to dialysis patients). The medical director is ultimately responsible for this decision; in some instances, short-term exposure to contaminants limited to one treatment may be the optimal choice, rather than not receiving the dialysis treatment at all. Refer to V182.

If the water supply utility has notified the city that the source water is highly chlorinated due to a water main break, flooding, or bacterial contamination of the municipal system, the dialysis facility must conduct more frequent monitoring of chlorine/chloramines (i.e., every 30-60 minutes).

For frequency of monitoring water for chemical contaminants, refer to V201 for Reverse Osmosis (RO) systems and to V206 for Deionization (DI) systems.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 37 (the full 469-page PDF) before quoting it.