V185 Easy access to meters, gauges and sample ports
Written V0185 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
8Environment (access to ports and meters)
The layout of the water purification system should provide easy access to all components of the system, including all meters, gauges, and sampling ports used for monitoring system performance.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004 8 Environment
An area for processing samples and performing on-site tests is also recommended.
Additional Guidance:
Occasionally, systems may be located in an area with limited space, Regardless of the layout of the water purification system, provision must be made to allow staff immediate access all equipment, ports, etc. to operate and monitor the system. In all cases, the operator should be able to describe and identify the various components and the distribution system.