V186 Critical alarms that sound in the treatment area

Written V0186 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 50 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

8Environment (alarms in the treatment area)

Critical alarms, such as those associated with deionizer exhaustion or low water levels in a storage tank, should be configured to sound in the patient treatment area, as well as in the water treatment room.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

Responsible staff members must be able to test the alarms to validate they can be heard in the treatment area. If alarms normally sound during certain events during the treatment day, documenting that these are heard in the treatment area will suffice for testing. Updated guidance from AAMI RD62:2001 recommend that alarms in the patient treatment area and water treatment rooms should be loud enough to be heard while patients are on dialysis, and should not be muted for more than 3 minutes.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 50 (the full 469-page PDF) before quoting it.