Middle Molecule Find

Rulebook › V-tags › V191

V191 Softeners: hardness testing and log

Written V0191 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 55 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40 (a)

6.2.4 Softeners: (Testing hardness/log)

Users should ensure that test accuracy and sensitivity are sufficient to satisfy the total hardness monitoring requirements of the reverse osmosis machine manufacturer. Total hardness of the water exiting the water softener should be measured at the end of each treatment day.

Water hardness test results should be recorded in a water softener log.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

6.2.4 Softeners

Softener monitoring, which should be performed daily, involves testing effluent water for total hardness to ensure that the limits established by the reverse osmosis machine manufacturer are not exceeded. In the case of automatically regenerating softeners, monitoring also includes verification that the brine tank contains a sufficient supply of undissolved sodium chloride and that the control valve timer, when present, indicates the correct time of day.

Testing for hardness should be performed using an ethylenediaminetetracetic acid (EDTA) titration test, with “dip and read” test strips, or a similar method.

The hardness test conducted at the end of the day will indicate the overall effectiveness of the water softener under worst-case conditions and ensure that the softener is sized correctly—that is, it has sufficient capacity, expressed in grains of calcium carbonate.

Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include: For softeners: monitor product water softness daily at the end of the treatment day for hardness as calcium carbonate, <l grain/gal, unless otherwise specified by the manufacturer of the reverse osmosis equipment.

Additional Guidance:

The water softener log may be incorporated as a part of another log kept for the water treatment system. For hardness tests requiring color differentiation, the person performing the analysis should be capable of correctly interpreting the results, including the ability to distinguish and accurately identify any colorimetric or other readings. If the person cannot differentiate these colors, an automated meter should be used.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 55 (the full 469-page PDF) before quoting it.