Rulebook › V-tags › V190

V190 Softeners: regeneration, timers and brine tank

Written V0190 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 54 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.4 Softeners: (auto regen/timers/salt/salt level)

6.2.4 Softeners (monitoring)

Prior to exhaustion, softeners should be restored; that is, new exchangeable sodium ions are placed on the resin by a process known as “regeneration,” which involves exposure of the resin bed to a saturated sodium chloride solution.

6.2.4 Softeners

Timers should be checked at the beginning of each day and should be interlocked with the RO system so that the RO is stopped when a softener regeneration cycle is initiated.

The softener brine tank should be monitored daily to ensure that a saturated salt solution exists in the brine tank. Salt pellets should fill at least half the tank. Salt designated as rock salt should not be used for softener regeneration since it is not refined and typically contains sediments and other impurities that may damage O-rings and pistons and clog orifices in the softener control head.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.2.4 Softeners

Water containing calcium or magnesium can form relatively hard deposits and is referred to as “hard water.” Water that has had these elements replaced by sodium ion exchange is called “soft water,” hence, the term “softener.” Softeners also remove other polyvalent cations, most notably iron and manganese, although they are somewhat limited in their effectiveness in this regard. The primary use of softeners in hemodialysis water systems is to prevent hard water deposits from damaging sensitive reverse osmosis membranes.

A softener is a cylinder or vessel that contains insoluble spheres or beads, called “resin,” to which sodium ions are attached. During operation, exchangeable sodium ions in the resin are progressively replaced by calcium and magnesium ions. When all the sodium ions have been used, the resin bed has reached a condition referred to as “exhaustion.” Softeners that automatically regenerate also include a brine tank, from which saturated sodium chloride solution is drawn during regeneration, and a control valve that regulates regeneration and service cycles.

Additional Guidance:

Softeners are not required in every facility; instead, the source water should determine the necessary water treatment components. If softeners are in use, the facility must follow these requirements.

Suggested monitoring guidelines from ANSI/AAMI RD52, Table 4 include: For the water softener regeneration cycle, monitor the regeneration cycle timer setting daily at the beginning of the day to ensure the softener timer is set to the correct time. Alternatively, verify that the timer is set by the facility to allow multiple tanks to backwash in sequence, rather than all at once.

Automatically regenerated water softeners shall be fitted with a mechanism to prevent water containing high concentrations of sodium chloride used during regeneration from entering the product water line during regeneration. The requirement to prevent water with a high concentration of sodium from entering the product water line is essential when a facility offers nocturnal dialysis, as regeneration cycles for most components are set for nighttime.

The face of the timers used to control the regeneration cycle should be visible to the user. The timer box cover should have a clear window allowing the timers to be seen, or the cover should be removed when timers need to be viewed. Facility policy should define the expected level of salt in the brine tank, with a minimum requirement that salt pellets fill at least half of the tank.

Rarely are dual bed softeners used. These allow one bed to function while the other bed regenerates. If used, the facility is responsible for the correct installation of the dual bed softeners to ensure that the effluent from the regenerating bed goes to drain, rather than downstream to other water treatment components, or that regeneration occurs when the RO is not in use.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 54 (the full 469-page PDF) before quoting it.