Middle Molecule Find

Rulebook › V-tags › V203

V203 Deionization: alarm and divert to drain

Written V0203 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 73 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.8 Deionization (alarms/divert to drain)

An audible and visual alarm shall be activated when the product water resistivity falls below this level and the product water stream shall be prevented from reaching any point of use, for example by being diverted to drain. The alarm must be audible in the patient care area. [ANSI/AAMI RD52:2004 guidance directs users to RD62:2001, Section 4.3.6]

ANSI/AAMI RD52:2004

5.2.8 Deionization

Specifically, the resistivity monitor following the final deionizer bed shall be connected to an audible and visible alarm in the dialysis treatment area, and the DI system shall divert product water to drain or otherwise prevent product water from entering the distribution system should an alarm condition occur. Under no circumstances shall DI be used when the product water of the final bed has a resistivity below 1 megohm-cm.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.2.8 Deionization

DI has a finite capacity that, when exceeded, will cause dangerously high levels of contaminants in the product water. Fortunately, the quality of product water from DI is easily monitored by resistivity monitors that, when used as specified (e.g., minimum resistivity 1 megohm-cm or greater), can prevent inadvertent operation of an exhausted deionization system.

Additional Guidance:

Except for home patients, there must be capability for either an automatic divert-to-drain system or other means to prevent product water from entering the distribution system for any DI system in use. DI has a very limited capacity that, when exceeded, will cause dangerously high levels of contaminants in the product water. Due to the increased risk to patient health and safety with the use of water from an exhausted DI tank, manually stopping the flow before the RO product water reaches the patient may not be as sufficient as the use of an automated protection system, such as a divert-to-drain valve.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 73 (the full 469-page PDF) before quoting it.