Middle Molecule Find

Rulebook › V-tags › V204

V204 Deionization: carbon before, ultrafilter after

Written V0204 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 74 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.2.8 Deionization (carbon adsorption and ultrafilter)

Systems that include deionizers as a component shall also contain carbon adsorption upstream of the deionizer to avoid formation of carcinogenic nitrosamines. In all instances, deionizers shall be followed by an ultrafilter or other bacteria- and endotoxin-reducing device to remove microbiological contaminants that may originate in the deionizer resin bed.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.2.8 Deionization

Refer to RD62:2001, 4.3.6 Deionization: Feed water for deionization systems should be pretreated with activated carbon adsorption, or a comparable alternative, to prevent nitrosamine formation.

In all instances, deionizers shall be followed by an ultrafilter or other bacteria- and endotoxin-reducing device to remove microbiological contaminants that may originate in the deionizer resin bed.

5.2.9 Ultrafiltration

Endotoxin-retentive ultrafilters should be placed in dialysis water systems in locations downstream of deionization, if deionization is the last process in a water treatment system.

Additional guidance:

Endotoxin-reducing devices and endotoxin-reducing ultrafilters may be used interchangeably in these applications. The typical range of micron filter size is .001

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 74 (the full 469-page PDF) before quoting it.