V208 Water storage and distribution: design for bacterial control

Written V0208 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 78 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.3 Water storage and distribution

5.3.1 General: Design

A water storage and distribution system should be designed specifically to facilitate bacterial control, including measures to prevent bacterial colonization and to allow for easy and frequent disinfection.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.3 Water storage and distribution

5.3.1 General

The function of the water storage and distribution system is to distribute product water from the purification cascade to its points of use, including individual hemodialysis machines, hemodialyzer reprocessing equipment, and concentrate preparation systems. A water storage and distribution system typically contains a large volume of water exposed to a large surface area of piping and storage tank walls. Because chlorine and chloramines are removed in the purification process, the water does not contain a bacteriostatic agent. This combination of circumstances predisposes wetted surfaces to bacterial proliferation and biofilm formation.

5.3.3 Water distribution systems

Two types of water distribution systems are used: direct feed systems and indirect feed systems. In a direct feed system, water flows directly from the last stage of the purification cascade to the points of use. In an indirect feed system, water flows from the end of the purification cascade to a storage tank. From there, it is distributed to the points of use. In general, direct feed systems offer the least favorable environment for bacterial proliferation. However, with a direct feed system the purification cascade must be sized to provide sufficient water to meet the peak demand, and the system must have sufficient pressure at the end of the purification cascade to distribute the water to the points of use. Those two requirements often preclude the use of a direct feed system.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.5.3 Water storage and distribution

Direct feed water distribution systems typically return unused water to the feed side of the reverse osmosis unit. If the pressure at the end of the distribution loop decreases to a value below the water pressure at the inlet to the reverse osmosis pressurizing pump, retrograde flow of non-purified water into the distribution loop can occur. To minimize this risk, the AAMI RDD Committee recommends that dual check valves be used to prevent retrograde flow and that the pressure at the end of the distribution loop be monitored.