V209 Water storage tanks: design and disinfection

Written V0209 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 79 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.3.2 Water storage: tank shape/configuration

When used, storage tanks should have a conical or bowl- shaped base and should drain from the lowest point of the base. Storage tanks should have a tight-fitting lid and be vented through a hydrophobic 0.2 µm air filter. The filter should be changed on a regular schedule according to the manufacturer’s instructions. A means shall be provided to effectively disinfect any storage tank installed in a water distribution system.

7.1 General strategies for bacterial control [in storage tanks]:

An ultrafilter, distal to the storage tank, or some other form of bacterial control device is recommended.

Storage tanks are therefore not recommended for use in dialysis systems unless they are frequently drained and adequately disinfected.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.3.2 Water storage

Internal spray mechanisms can facilitate effective disinfection and rinsing of a storage tank.

7 Strategies for bacterial control

7.1 General

A storage tank in the distribution system greatly increases the volume of fluid and surface area available and can serve as a niche for water bacteria. It may be necessary for the user to scrub the sides of the tank to remove bacterial biofilm if the tank design and maintenance are not adequate to prevent bacterial proliferation.

Additional Guidance:

If existing facilities with older storage tanks can demonstrate a history of water and dialysate cultures being below action levels, replacement of the existing tanks is not required. AAMI recommendation at 7.1 notes that the frequency of disinfection will vary with the design of the system, but disinfection should be performed monthly. In order to demonstrate a history of effective disinfection procedures, the facility is expected to demonstrate acceptable results for the most recent 3 months of water and dialysate cultures.

Bacterial control device(s) in use following the storage tank may include an ultrafilter or individual filters in the water supply line in each patient’s dialysis machine.

If manufacturer instructions contain disinfection procedures for the storage tank, this should be followed. Tanks which fill from the top and drain from the bottom may, in fact, drain several times a day. The goal is to not have stagnant water. A properly designed and functional storage tank replaces its total volume throughout the day as part of the normal operation and does not require manual or frequent draining of the tank.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 79 (the full 469-page PDF) before quoting it.