V210 Water storage tanks: bacteria and endotoxin monitoring

Written V0210 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 80 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

6.3 Water storage and distribution

6.3.2 Water storage (monitoring)

Routine monitoring of water storage tanks for bacteria and endotoxin levels is generally accomplished indirectly by monitoring the water at the first outlet to the distribution loop (see 6.3.3). If direct monitoring of a water storage tank is performed as part of a troubleshooting process, bacteria and endotoxin levels shall be measured as specified in ANSI/AAMI RD62:2001 (see 2.3). All bacteria and endotoxin results should be recorded on a log sheet.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

Additional Guidance:

Suggested water storage tank monitoring guidelines from ANSI/AAMI RD52, Table 4 include:
Measure bacterial growth and pyrogens, weekly, until a pattern of consistent compliance can be demonstrated. Action levels for bacterial growth are 50 CFU/mL; for endotoxin 1 EU/mL. Expect action to be taken for levels above 50 CFU/mL and above 1 EU/mL.

A “pattern of consistent compliance” could be demonstrated by showing results within these limits on weekly cultures for at least four weeks in a row. Laboratory generated reports are an acceptable alternative to recording results on a log if the laboratory provides an aggregate report allowing multiple monthly reports to be easily compared for trends.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 80 (the full 469-page PDF) before quoting it.