V215 UV irradiators followed by endotoxin reduction
Written V0215 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.3.4.1 Ultraviolet irradiators (filter)
UV irradiators should be followed by a means of reducing endotoxin concentrations, such as an ultrafilter in the purified water distribution system or reverse osmosis in the pretreatment cascade.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.3.4.1 Ultraviolet irradiators
Ultraviolet irradiation also can be used to control bacteria in the pretreatment section of a water purification system, such as following carbon adsorption beds to reduce the bacterial burden presented to a reverse osmosis unit. Because using UV irradiation to kill bacteria increases the level of endotoxins in the water, UV irradiators should be followed by a means of reducing endotoxin concentrations, such as an ultrafilter in the purified water system or reverse osmosis in the pretreatment cascade.
Additional Guidance:
An ultrafilter or other endotoxin-retentive device may be used. If an RO follows the UV, the RO functions as an ultrafilter. This requirement is not intended to apply to UV irradiators placed in some bicarbonate distribution systems.
https://middlemolecule.com/rulebook/tags/v215/