V223 Concentrate preparation: compatible materials

Written V0223 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 94 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.4 Concentrate preparation

5.4.2 Materials compatibility

All components used in concentrate preparation systems (including mixing and storage tanks, pumps, valves, and piping) shall be fabricated from materials (e.g., plastics or appropriate stainless steel) that do not interact chemically or physically with the concentrate so as to affect its purity, or with the germicides or germicidal procedure used to disinfect the equipment. The use of materials that are known to cause toxicity in hemodialysis, such as copper, brass, galvanized material, and aluminum, are specifically prohibited.

CMS's interpretive guidance

No separate guidance under this tag.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 94 (the full 469-page PDF) before quoting it.