V222 Acid concentrate bulk storage tanks
Written V0222 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.4 Concentrate preparation
5.4.3 Bulk storage tanks (acid concentrate)
Procedures should be in place to control the transfer of the acid concentrate from the delivery container to the storage tank to prevent the inadvertent mixing of different concentrate formulations. If possible, the tank and associated plumbing should form an integral system to prevent contamination of the acid concentrate. The storage tanks and inlet and outlet connections, if remote from the tank, should be secure and labeled clearly.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.4 Concentrate preparation
5.4.1 General
Acid concentrates supplied in 55-gallon drums or gallon containers by the manufacturer are the responsibility of that manufacturer (see AAMI 2.4). In some cases, the manufacturer will pump the acid concentrate from the 55-gallon drums into a holding tank at the dialysis facility. If the acid concentrate is pumped into a bulk storage tank at the dialysis facility, the user is responsible for maintaining the concentrate in its original state and to ensure that the correct formula is used according to the patient’s prescription.
Additional Guidance:
Acid concentrates supplied in 55 gallon drums or gallon containers are the responsibility of the manufacturer until delivered; proper handling after delivery is the responsibility of the dialysis facility. If an inlet opening to the acid concentrate storage system is located on the outside of the building, the inlets should be secure and labeled.