V227 Self-designed mixing systems: lab verification

Written V0227 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 96 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

6.4.1 Mixing systems (self-designed)

If a facility designs its own system, procedures should be developed and demonstrated to ensure proper mixing of the concentrate, including establishment of acceptable limits for tests of proper concentration. The adequacy of those procedures must be verified using an independent laboratory that is capable of meeting the requirements of ANSI/AAMI RD61:2000 (see 2.4). Verification can be accomplished by testing a sample from each batch prepared over a 3-day period.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

If a facility designs its own mixing systems, there must be documentation of the verification testing available for review.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 96 (the full 469-page PDF) before quoting it.