Middle Molecule Find

Rulebook › V-tags › V228

V228 Labeling mixing tanks, storage tanks and jugs

Written V0228 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 96 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40 (a)

5.4.4.1 Mixing systems (labeling)

Labeling strategies should permit positive identification by anyone using the contents of mixing tanks, bulk storage/dispensing tanks, and small containers intended for use with a single hemodialysis machine.

Mixing tanks: Prior to batch preparation, a label should be affixed to the mixing tank that includes the date of preparation and the chemical composition or formulation of the concentrate being prepared. This labeling should remain on the mixing tank until the tank has been emptied.

Bulk storage/dispensing tanks: These tanks should be permanently labeled to identify the chemical composition or formulation of their contents.

Concentrate jugs: At a minimum, concentrate jugs should be labeled with sufficient information to differentiate the contents from other concentrate formulations used at the facility.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.4.4.1 Mixing systems

Requirements for [labeling strategies that permit] such positive identification will vary among facilities, depending on the differences between concentrate formulations used and on whether single or multiple dialysate proportioning ratios are used. (The use of multiple dialysate proportioning ratios [e.g., 35X and 45X] in a single facility is strongly discouraged.)

Mixing tanks: Using a photocopy of the concentrate manufacturer’s package label provides a convenient and comprehensive means of identifying the chemical composition or formulation of the concentrate; however, the lot number and expiration date should be marked out because they apply only to the dry powder.

As with mixing tanks, bulk storage/dispensing tank labeling can be conveniently accomplished by affixing a copy of the concentrate manufacturer’s package label.

Concentrate jugs are typically non-disposable vessels provided by hemodialysis machine manufacturers and having a capacity sufficient for one or two hemodialysis sessions. The extent of labeling for these containers depends on the variety of concentrate formulations used and on whether the facility uses dialysis machines with different proportioning ratios (a practice that is strongly discouraged).

Additional Guidance:

If multiple dialysate proportioning ratios are in use, applicable staff and the medical director should be able to describe safeguards in place to prevent mismatching dialysate components/machines. There should be no incidents of ratio mismatch, for example, 35X acid used with a machine set for 45X. Labels made by the facility are acceptable as long as the required information is included. Labels should be used to alert staff when bleach or ozone is in a tank or concentrate jug during disinfection. If a group of jugs are being disinfected at once, a process control (such as a label or sign marking the area in use) could be used rather than individual labels for each jug.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 96 (the full 469-page PDF) before quoting it.