V232 Bicarbonate concentrate mixing tanks

Written V0232 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 100 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.4.4.3 Bicarbonate concentrate mixing systems: empty/ disinfect/prevent corrosion

Bicarbonate concentrate mixing tanks should be designed to drain completely[.]

Mixing tanks should have a tight-fitting lid and should be designed to allow all internal surfaces to be disinfected and rinsed.

Because concentrate solutions are highly corrosive, mixing systems should be designed and maintained to prevent corrosion.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.4.4.3 Bicarbonate concentrate mixing systems

Bicarbonate concentrate mixing tanks should be designed to drain completely; for example, [bicarbonate mixing tanks] should have a sloping bottom and a drain at the lowest point. High- and low-level alarms can prevent overfilling and air damage to the pump.

A translucent tank allows users to see the liquid level; the use of sight tubes is not recommended because of the potential for microbial growth, such as bacteria, algae, and fungi.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.5.4.4.3 Bicarbonate concentrate mixing systems

Bicarbonate concentrates have been shown to support bacterial growth and to provide another source of initial bioburden capable of rapidly increasing after dilution. . . Therefore, additional precautions should be taken when preparing and handling bicarbonate concentrate to avoid excess growth of haloduric organisms. Also, prompt use of bicarbonate concentrates prepared in dialysis facilities from powder and purified water is strongly recommended.

Additional Guidance:

High and low-level alarms are not required. Facilities should maintain the system clean and free of spills of concentrate to reduce the potential of corrosion. Expect that systems will have little to no build-up of precipitate.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 100 (the full 469-page PDF) before quoting it.