Middle Molecule Find

Rulebook › V-tags › V231

V231 Acid concentrate mixing tanks

Written V0231 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 99 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.4.4.2 Acid concentrate mixing systems: empty completely/prevent corrosion

Acid concentrate mixing tanks should be designed to allow the inside of the tank to be completely emptied and rinsed according to the manufacturer’s instructions when concentrate formulas are changed.

Acid concentrate mixing tanks should be emptied completely before mixing another batch of concentrate.

Because concentrate solutions are highly corrosive, mixing systems should be designed and maintained to prevent corrosion.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.4.4.2 Acid concentrate mixing systems

Use of a tank with a sloping bottom that drains from the lowest point is one means of facilitating this process.

Additional Guidance:

Facility policy and practice must ensure the tank is completely emptied between mixing batches of concentrate.

Facilities should maintain the system clean and free of spills of concentrate to reduce the potential of corrosion.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 99 (the full 469-page PDF) before quoting it.