Middle Molecule Find

Rulebook › V-tags › V237

V237 Concentrate distribution: compatible materials

Written V0237 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 104 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.5 Concentrate distribution:

5.5.1 Materials compatibility

All components used in concentrate distribution systems (including concentrate jugs, storage tanks, and piping) that contact the fluid shall be fabricated from nonreactive materials (e.g., plastics or appropriate stainless steel) that do not interact chemically or physically with the concentrate so as to affect its purity. The use of materials that are known to cause toxicity in hemodialysis, such as copper, brass, galvanized material, and aluminum, are specifically prohibited.

CMS's interpretive guidance

No separate guidance under this tag.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 104 (the full 469-page PDF) before quoting it.