V238 Elevated bicarbonate distribution tanks

Written V0238 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 104 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.5.2 System configurations: elevated tanks

Elevated tanks for bicarbonate concentrate distribution should be equipped with conical or bowl-shaped bottoms, tight-fitting lids, a spray mechanism, and high- and low-level alarms. Any air vents should have 0.2 µm hydrophobic vent filters.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.5.2 System configurations

Concentrate may be distributed from a central preparation point using reusable concentrate jugs that contain sufficient concentrate for one to two treatments, or it may be distributed through a piping system that provides concentrate connections at each treatment station. A combination of these two systems may also be used, with some concentrates distributed by concentrate jug and others through a piping system. Two common configurations used for distributing concentrate through a piping system are gravity feed and pressurized. Gravity feed systems require an elevated tank; pressurized systems deliver the concentrate using a pump and motor and do not require an elevated tank. The maximum allowable concentrate delivery pressure is specified by the manufacturer of the dialysate delivery machine and should not be exceeded.

Additional Guidance:

System configuration will vary based on the facility dynamic. When elevated tanks are used, they are usually smaller than those used for preparing concentrates.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 104 (the full 469-page PDF) before quoting it.