Written V0240 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 106 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.5.4 Bicarbonate concentrate distribution systems (Ultraviolet Irradiation)

UV irradiation devices that are used to control bacteria proliferation in the pipes of bicarbonate concentrate distribution systems should be fitted with a low-pressure mercury lamp that emits light at a wavelength of 254 nm and provides a dose of radiant energy of 30 milliwatt- sec/cm2. The device should be sized for the maximum anticipated flow rate according to the manufacturer’s instructions and be equipped with an on-line monitor of radiant energy output that activates a visual alarm indicating that the lamp should be replaced.

Alternatively, the lamp should be replaced on a predetermined schedule according to the manufacturer’s instructions to maintain the recommended radiant energy output. Disinfection of the bicarbonate concentrate distribution system should continue to be performed routinely.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

This requirement applies to UV irradiators used in the bicarbonate delivery system, rather than to an UV irradiator used in the water treatment/distribution system. A facility may have an UV irradiator in one system, and not in the other. The UV irradiator in the bicarbonate delivery system may also be used to help breakdown the ozone used to disinfect the bicarbonate delivery system.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 106 (the full 469-page PDF) before quoting it.