V239 Bicarbonate distribution: disinfection

Written V0239 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 105 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.5.4 Bicarbonate concentrate distribution systems: weekly disinfection/dwell times/conc [B]icarbonate concentrate delivery systems should be disinfected on a regular basis to ensure that the dialysate routinely achieves the level of bacteriological purity [required by these standards].

For piped distribution systems, the entire system, including patient station ports, should be purged of bicarbonate concentrate before disinfection. Each patient station port should be opened and flushed with disinfectant and then rinsed; otherwise, it would be a “dead leg” in the system.

Appropriate dwell times and concentrations should be used as recommended by the manufacturer of the concentrate system. If this information is not available, bleach may be used at a dilution of 1:100 and proprietary disinfectants at the concentration recommended by the manufacturer for disinfecting piping systems.

6.5 Concentrate distribution:

It is recommended that the interval between disinfections not exceed 1 week. If the manufacturer does not supply disinfection procedures, the user must develop and validate a disinfection protocol.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.5.4 Bicarbonate concentrate distribution systems

Bicarbonate concentrates provide excellent media for bacterial proliferation. The manufacturer’s instructions can provide an initial disinfection schedule. However, this schedule may need to be adjusted on the basis of the user’s bacteriological monitoring.

All chemical disinfectants (e.g., bleach and peracetic acid products) that are compatible with dialysis machines can be used to disinfect bicarbonate concentrate delivery systems. However, some disinfectants attack biofilm better than others.

In the event that precipitation or salt build-up impedes flow through a piping system, cleaning with a 1:34 solution of 5% acetic acid (e.g., distilled white vinegar) is recommended. Some manufacturers supply bicarbonate concentrate systems with UV irradiation or ozone systems for bacterial control.

Additional Guidance:

Alternatively, a 5% citric acid solution may be used instead of acetic acid if the manufacturer allows. It is not expected that concentrates would be cultured or tested for endotoxin levels. Bicarbonate concentrates are monitored via dialysate cultures.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 105 (the full 469-page PDF) before quoting it.