V247 Concentrate outlets: separate, labeled, checked
Written V0247 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
5.5.5 Concentrate outlets (designated outlets and labeling)
6.5 Concentrate distribution
5.5.5 Concentrate outlets: separate/labeled/connection safety
To prevent mix-ups with delivery of two or more types of acid concentrate, each concentrate should have its own outlet. Concentrate outlets should be compatible with the dialysis machine and have a means of minimizing the risk that the wrong concentrate will be connected to an outlet. The dispensing outlets should be labeled with the appropriate symbol (see AAMI Table 3) indicating the proportioning ratio for the dialysis machine and should be color-coded blue for bicarbonate, red for acid.
6.5 Concentrate distribution
A daily check to ensure that the appropriate acid and bicarbonate concentrate is connected to the corresponding concentrate delivery line is recommended if the storage tank is not permanently connected to its distribution piping.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
ANSI/AAMI RD52:2004
5.5.5 Concentrate outlets
For piped concentrate distribution systems, each treatment station is equipped with a concentrate outlet for bicarbonate, one or more outlets for acid concentrate, and a product water outlet for connection to the inlet line of the dialysis machine.