V248 Concentrates matched to the machine's ratio

Written V0248 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 110 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

5.6 Dialysate proportioning [T]he acid and bicarbonate concentrates [must] be matched with respect to the proportioning ratio and with the model and setup configuration of the dialysis machine.
Several types of three-stream concentrates are available, with different ratios of acid concentrate to bicarbonate concentrate to water (see Table 3). The different proportioning types are not compatible with one another.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

5.6 Dialysate proportioning

Essentially, all dialysate is produced with three fluid streams: water, acid concentrate, and bicarbonate concentrate. This three-stream combination produces a highly buffered dialysate with a pH between 6.9 and 7.6. Dialysate can also be prepared from a single concentrate that contains acetate to provide a dialysate in which buffer is provided to the patient in the form of acetate, which is subsequently metabolized to yield bicarbonate. However, acetate containing dialysate is now rarely used in clinical practice.

Different manufacturers of dialysis machines use different methods of controlling the proportions of the concentrates. These can be generally grouped into two categories: “fixed proportioning” and “servo control.” With both methods, the operator can select a desired sodium and bicarbonate level, and the machine will make the necessary adjustments to achieve the selected levels. Both types use a redundant system of controls and monitoring. With fixed proportioning systems, the pumps are set to established volumes, and the final conductivity is verified. With servo control machines, the individual concentrates are added until the conductivity achieves the expected value. A final redundant conductivity monitor verifies the mixture. Some machines may also monitor the pH of the dialysate as an additional safeguard against gross errors in dialysate formulation.

Generally, bicarbonate is available in one or two forms for each proportioning type (in liquid, cartridge, or dry powder, and in various sizes). Each proportioning type has numerous acid concentrate formulations (“codes”) with different amounts of potassium, calcium, and magnesium ions, plus dextrose. To help differentiate between concentrates of different proportioning types, AAMI recommends that the manufacturer include a geometric symbol on the labels along with acid/base color coding.

Table 3- Symbols and color coding for different concentrate proportioning ratios
Concentrate typeAcid proportioning ratio (Red color coding)Geometric symbolBicarbonate Concentrate (Blue color coding)Comments
35X1:34SquareDry, liquid or cartridge
36.83X1:35.83CircleDry or liquidBicarbonate concentrate contains some NaCl
45X1:44TriangleDry, liquid or cartridge
36.1X1:35.1HexagonCartridgePower cartridges may be used for other proportioning ratios, except for 36.83X, in which the bicarbonate concentrate also contains NaCl

Typed by hand from pages 111-112 of Appendix H, because the PDF's table doesn't extract cleanly. Open page 111 (the full PDF).

Note 1 – The acid proportioning ratio refers to acid concentrate, water + bicarbonate concentrate

Note 2 – Acetate- containing concentrate is color -coded white

AAMI Rationale for the Development and Provision of This Recommended Practice

A.5.6 Dialysate proportioning

Dialysate is usually prepared by a proportioning system that sequentially adds acid concentrate and bicarbonate concentrate to purified water. These systems produce a buffered physiologic dialysate with a pH between 6.9 and 7.6. More recently, systems have been developed that use three concentrates (bicarbonate, sodium chloride, and an acid concentrate containing the remaining electrolytes) to allow more sophisticated variation of the dialysate composition during dialysis.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 110 (the full 469-page PDF) before quoting it.