V254 Samples taken before disinfection

Written V0254 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 119 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

7.2 Microbial monitoring methods

7.2.1 General

Samples should always be collected before sanitization/disinfection of the water treatment system and dialysis machines.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

Samples must be collected before sanitation and disinfection procedures begin, since these procedures can impact and interfere with the results of the microbial tests. If repeat cultures are performed after sanitation and disinfection, the system should be flushed completely before collecting samples.

If the microbial testing is performed by an outside laboratory, logistics for the timely receipt of evaluating culture results should be considered, i.e., sample collection time, transit time, receipt of sample during lab operating days/hours, etc. Samples sent to labs must arrive during the lab’s hours of operation. If disinfection is scheduled for Sunday, for example, samples may need to be collected on Wednesday or Thursday to allow transit time and arrival before the lab closes on Friday.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 119 (the full 469-page PDF) before quoting it.