V255 Repeat cultures after high counts
Written V0255 on the CMS-2567 and in QCOR.
Cited on this tag? What happens after a citation · One-page sheet for the binder
The regulation
ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)
7.2 Microbial monitoring methods
7.2.1 General (repeat cultures)
Cultures should be repeated when bacterial counts exceed the allowable levels. If culture growth exceeds permissible standards, the water system and dialysis machines should be cultured weekly until acceptable results are obtained. Additional samples should be collected when there is a clinical indication of a pyrogenic reaction or septicemia, and following a specific request by the clinician or the infection control practitioner.
If repeat cultures are performed after the system has been disinfected (e.g., with formaldehyde, hydrogen peroxide, chlorine, or peracetic acid), the system should be flushed completely before collecting samples. Drain and flush storage tanks and the distribution system until residual disinfectant is no longer detected before collecting samples.
CMS's interpretive guidance
Interpretive Guidance § 494.40(a)
Responsible staff members must be able to describe what is done when bacterial counts exceed action levels.
Connected tags
Tags whose guidance points here
https://middlemolecule.com/rulebook/tags/v255/