V257 Plate counts: sample timing, no calibrated loop

Written V0257 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 121 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

7.2.3 Heterotrophic plate count

Samples that cannot be cultured within 1 to 2 hours can be refrigerated for up to 24 hours.

Use of a calibrated loop to apply the sample to the agar plate is NOT permitted.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

7.2.3 Heterotrophic plate count

The reference method for culturing is the membrane filtration technique. With this method, a known volume of sample or diluted sample is filtered through a 0.45 µm membrane filter and the membrane filter is aseptically transferred to the surface of an agar plate. Trypticase soy agar (TSA, a soybean casein digest agar) is the medium of choice for culturing water and dialysate; other acceptable media include standard methods agar and plate count agar (also known as TGYE). Blood and chocolate agars are not appropriate for this test. The spread plate technique may also be used. With this method, an inoculum of at least 0.5 mL of sample is spread equally over the surface of the agar plate.

AAMI Rationale for the Development and Provision of This Recommended Practice

A.7 Strategies for bacterial control

A.7.2.3 Heterotrophic plate count

Sensitive culturing methods must be used to measure the low total viable microbial counts permitted for water used for hemodialysis applications under the provisions of ANSI/AAMI RD62:2001 and recommended for dialysate in this recommended practice. The membrane filter technique is particularly suited for this application because it permits large volumes of water to be assayed. Because the membrane filter technique may not be readily available in clinical laboratories, the spread plate assay can be used as an alternative.

However, if the spread plate assay is used, a calibrated loop shall not be used to apply sample to the plate. The sensitivity of an assay performed with a calibrated loop is low. A standard calibrated loop transfers 0.001 mL of sample to the culture medium, so that the minimum sensitivity of the assay is 1,000 CFU/mL. This sensitivity is unacceptable when the maximum allowable limit for microorganisms is 200 CFU/mL. Therefore, when the spread plate method is used, a pipette should be used to place 0.1 mL to 0.5 mL of water on the culture medium.

Additional Guidance:

Responsible staff must be knowledgeable of the culture methods in use and must inform any outside labs that the samples are water and dialysate and that use of a calibrated loop to inoculate the culture plate is not acceptable.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 121 (the full 469-page PDF) before quoting it.