Middle Molecule Find

Rulebook › V-tags › V256

V256 Dip samplers: quality program and incubation

Written V0256 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD52:2004, adopted by reference at 42 CFR 494.40(a) · Appendix H Rev. 246, page 120 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD52:2004 Requirements as Adopted by Reference 42 CFR 494.40(a)

7.2.3 Heterotrophic plate count (dip samplers)

Dip samplers may be used for bacterial surveillance… in conjunction with a quality assurance program designed to ensure their appropriate use. Elements of the quality assurance program should include staff training in areas such as the correct methods of inoculation, incubation, and interpretation, and verification involving duplicate samples sent to a certified laboratory on at least an annual basis. Plates shall be incubated at 35 °C for 48 hours.

Colonies should be counted using a magnifying device.

CMS's interpretive guidance

Interpretive Guidance § 494.40(a)

ANSI/AAMI RD52:2004

7.2.3 Heterotrophic plate count

This method is an indicator of water quality only and is not to be confused with total heterotrophic plate counts, which require much longer incubation times at 28° C.

If a more accurate count from plates containing fewer than 30 or more than 300 colonies is desired, larger or smaller volumes may be cultured. Smaller volumes can be obtained by making 1:10 serial dilutions in sterile phosphate buffer. If larger volumes are required, the membrane filtration method should generally be used.

Erratic colony counts . . . may indicate the presence of biofilm since sloughing of biofilm may occur with release of bacteria into the water. [W]hen contamination persists in spite of frequent and aggressive disinfection, it may be necessary to determine if biofilm is present in the system.

Additional Guidance:

Facilities that use dip samplers must send duplicate samples to a laboratory at least annually to evaluate the accuracy of the testing done with the dip samplers.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 120 (the full 469-page PDF) before quoting it.

Connected tags

Tags whose guidance points here