Middle Molecule Find

Rulebook › V-tags › V312

V312 Informing patients about reuse

Written V0312 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 137 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

6.2 Informed consent

All patients in a dialysis facility will be fully informed regarding reuse of dialyzers. Printed material such as brochures describing the facility’s services should contain a statement about dialyzer reprocessing if reuse is performed.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.6.2 Informed consent

Establishing QA practices such as those recommended here and sharing information with patients may aid in solutions to these problems.

The National Kidney Foundation and the American Association of Kidney Patients recommend that patient consent for dialyzer reuse be obtained.

Additional Guidance:

CMS does not require specific written patient consent, but does require that patients be informed that the facility does reprocess dialyzers and be informed about that process. Refer to V460 under the Condition for Patients’ rights.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 137 (the full 469-page PDF) before quoting it.