V313 Reprocessing equipment: design, testing, QC

Written V0313 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 138 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

7 Equipment

Each piece of equipment used for reprocessing shall be appropriately designed, constructed, and tested to perform its intended task. Satisfactory operation of manual and automated systems shall be ensured by appropriate functional tests. Strict QC and QA shall be maintained for any type of dialyzer reprocessing equipment. Additionally, complete documentation of system function, operating procedures, potential system failures, and dialyzer-reuse criteria shall be included in the dialyzer reprocessing manual, known to the operator, and available for review.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.7 Equipment

Types of reprocessing systems vary from sophisticated microprocessor-controlled systems to hand-operated valving systems.

It is particularly important that appropriate equipment is used properly to ensure all water that comes into contact with the fluid pathways for blood or dialysate is of AAMI quality, because the blood side of the dialyzer might take up endotoxins that could be released into the circulation system during the subsequent dialysis.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 138 (the full 469-page PDF) before quoting it.