V317 Repairs to the reprocessing system

Written V0317 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 141 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

7.2.4 Repairs

If the reprocessing system fails to function as expected, qualified personnel should investigate and repair the problem. The reprocessing system function testing should be repeated after repairs of automated equipment and, if appropriate, after repairs of manual equipment before either the dialyzer is reprocessed or the reprocessed dialyzer is used for clinical dialysis.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

There should be documentation to verify that testing of repaired equipment for expected functioning was completed successfully prior to returning the equipment to service.