Middle Molecule Find

Rulebook › V-tags › V318

V318 Reprocessing area and ventilation

Written V0318 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 142 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50 (b)(1)

8 Physical plant and environmental safety considerations

8.1 Reprocessing area and ventilation

The reprocessing area should be designed to suit the operation carried out and maintain acceptable ambient concentrations of harmful substances (see Table 1). The area should be kept clean and sanitary. It may be part of the dialysis treatment area, as long as equipment used is properly designed and vented to meet the requirements for environmental safety (see [AAMI] 8.5).

Table 1—OSHA environmental exposure limits (29 CFR 1910, 1 July 1998), except as indicated
Substance/materialLimits (PEL)a
Acetic acid10 ppm TWAb
Chlorine dioxide (syn: chlorine oxide)0.1 ppm TWA
Citric acidNone developed
Formaldehyde0.75 ppm TWA
2 ppm STELc(15 min)
0.5 ppm action level
Glutaraldehyde0.2 ppm ceiling NIOSH/OSHA
Hydrogen peroxide1 ppm TWA
Peracetic acidNone developed
Phenol5 ppm TWA

Typed by hand from page 142 of Appendix H, because the PDF's table doesn't extract cleanly. Open page 142 (the full PDF).

ppm = parts per million
a ) PEL (permissible exposure limit) represents the limit of what employees can be exposed to; PELs can be TWAs or STELs.
b ) TWA (time-weighted average) represents the limit of what an employee can be exposed to in an eight-hour period.
c ) STEL (short-term exposure limit) represents the limit of what an employee can be exposed to in any 15-minute time period.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

Although the equipment for air testing may not be kept on-site, it should be available for use if staff or patients complain about germicide vapors.

There should be a schedule for routine air-level testing for germicides vapors, along with references describing the safe exposure levels, and if there are any circumstances which would require an unscheduled test.

The reprocessing area must be kept clean and free of clutter. Blood splashes should be immediately cleaned and the affected surfaces disinfected.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 142 (the full 469-page PDF) before quoting it.