Middle Molecule Find

Rulebook › V-tags › V321

V321 Storage of dialyzers and reprocessing materials

Written V0321 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 144 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

8.2 Storage area

Reprocessing materials, hemodialyzers awaiting reprocessing, and reprocessed hemodialyzers should be stored so as to minimize deterioration, contamination, or breakage. New, used, and reprocessed dialyzers should be segregated to make clear the status of each group of dialyzers. Environmental contamination of the storage area should be controlled and monitored, if the personnel determine those actions to be necessary. Storage areas for new dialyzers and reprocessing materials should be designed to facilitate rotation of stock and cleaning. Storage arrangements should also take into account fire safety considerations, OSHA regulations, and other appropriate regulations.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

“Clean” and “dirty” dialyzers must be stored separately; the reprocessing cycle status of any dialyzer must be clearly apparent at all times. Stock must be organized to allow rotation and prevent use of out-of-date materials. Reprocessed dialyzers in storage should be protected from unauthorized access to prevent tampering and any potential damage from manipulation by untrained users.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 144 (the full 469-page PDF) before quoting it.