Written V0322 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 145 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

9 Reprocessing supplies

9.1 Specifications and testing

Each reprocessing material should meet a written specification. The fulfillment of that requirement may be determined by certification by the product’s supplier that the product meets necessary specifications, labeling for its intended purpose, or by testing procedures by trained personnel, as appropriate. The requirement may also be complied with by purchasing a specific grade as specified by the process, such as USP citric acid. When the user performs testing, he or she should maintain a log of the date of test, the identifying number (lot number) of the batch, the person performing any testing, and the test results.

When bleach is purchased from a commercial outlet, the labeled concentration should be between 5.25% and 6.15%, and the formula should not contain fragrances or scents.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

ANSI/AAMI RD47:2002/A1:2003 9 Reprocessing supplies

9.1 Specifications and testing

Over the past few years, bleach (sodium hypochlorite) manufacturers have begun selling household bleach in many new formulas. The concentration of sodium hypochlorite has gone from 5.25% to 6.15% in many cases. The CDC has not changed its recommendations for diluting the bleach to take into account these percentage changes.

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.9 Reprocessing supplies

A.9.1 Specifications and testing

Testing of all incoming materials had been proposed. In recognition of the fact that most medical supplies are certified by the vendor and not tested by the user, the AAMI RDD Committee decided to recommend that supplies need not be tested by the facility doing hemodialyzer reprocessing if they are marketed for hemodialyzer reprocessing.

Additional Guidance:

Most materials used in reprocessing will not require testing by the user since they are validated and certified by a vendor. If the facility is using a material not commonly used in reprocessing, it may not be certified by a supplier as meeting necessary specifications . For these materials, there must be documentation of testing done by trained staff members to verify that the product meets the necessary specifications.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 145 (the full 469-page PDF) before quoting it.