V323 Supplies: first in, first out

Written V0323 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 146 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

9.2 Inventory control

Reprocessing supplies should be used on a first-in, first-out basis, and outdated supplies should be identified and discarded.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.9.2 Inventory control

The AAMI RDD Committee suggested that supplies should be used on a first-in, first-out basis to avoid deterioration over time in storage.

Additional Guidance:

Expired supplies must be discarded or quarantined for return to the vendor. Note that testing strips (which may be found in multiple locations in the reuse room and treatment floor) may require dating when opening and should be discarded based on the number of days since opening. Note that some reprocessing supplies, such as blood port caps, do not have expiration dates.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 146 (the full 469-page PDF) before quoting it.