Rulebook › V-tags › V331

V331 Transporting and holding used dialyzers

Written V0331 on the CMS-2567 and in QCOR.

Enforces ANSI/AAMI RD47:2002/A1:2003, adopted by reference at 42 CFR 494.50(b)(1) · Appendix H Rev. 246, page 151 · checked 12 Nov 2026

Cited on this tag? What happens after a citation · One-page sheet for the binder

The regulation

ANSI/AAMI RD47:2002/A1:2003 Requirements as Adopted by Reference 42 CFR § 494.50(b)(1)

11 Reprocessing

11.1 Transportation and handling

Persons handling used dialyzers during transportation shall do so in a clean and sanitary manner maintaining Standard Precautions until the dialyzer is disinfected both internally and externally. To inhibit bacterial growth, dialyzers that cannot be reprocessed within 2 hours should be refrigerated and not allowed to freeze. Other transportation and handling issues (such as prolonged delays in reprocessing) not described in this recommended practice shall be validated and documented by the responsible party.

Read 42 CFR 494.50 on the eCFR

CMS's interpretive guidance

Interpretive Guidance § 494.50(b)(1)

AAMI Rationale for the Development and Provisions of this Recommended Practice

A.11.1 Transportation and handling

During the 2002 revision of this recommended practice, the AAMI RDD Committee recognized that the refrigeration temperature of the dialyzers stored for extended periods of time was not specified. It was decided to recommend that dialyzers not reprocessed within 2 hours should be refrigerated and not allowed to freeze. The AAMI RDD Committee believed that this was sufficient to retard bacterial growth.

Additional Guidance:

Personnel should wear appropriate PPE to handle used dialyzers until reprocessing is complete. The infection control recommendations from the Centers for Disease Control and Prevention for transporting dialyzers include the following: "For dialyzers and blood tubing that will be reprocessed, cap dialyzer ports and clamp tubing. Place all used dialyzers and tubing in leakproof containers for transport from station to reprocessing or disposal area."

This requirement for appropriate transportation and handling procedures for used dialyzers would be considered met if the tubing is disposed of at the patient’s chairside directly into a waste receptacle and all the ports on the dialyzer are immediately capped.

All dialyzer ports should be capped when the dialyzer is not in use or not being currently reprocessed, to prevent spills of blood or blood products, leakage of germicide, and entrance of air into the dialyzer. If used dialyzers are transported in a common carrier (e.g., a basket) the potential for cross-contamination must be eliminated (i.e. exteriors must be free of visible blood and all ports capped or each dialyzer contained in a sealed bag).

If used dialyzers are refrigerated prior to reprocessing, the facility's policy must define maximum refrigeration times, temperature ranges, and quality controls in place to ensure the practice is safe.

Practices such as allowing dialyzers to remain at room temperature for prolonged periods during the reprocessing process (e.g., after rinsing and before filling the dialyzer with germicide) must be validated to ensure that patient safety is not compromised.

If dialyzers are sent to an off-site location for reprocessing, the survey process must include a visit to that site to determine compliance with this and all other reprocessing requirements.

Verbatim from Appendix H Rev. 246, with CMS's headings, lists and bold. Check page 151 (the full 469-page PDF) before quoting it.